Readme_bowhound The bowhound system was an underway pCO2 system that was installed at the seachest. The prototype system used a low cost Li-COR Gas Hound and a Liqui Cell membrane extractor ("beer gasser") which was used to equilibrate headspace and water. The unit was not temperature controlled and operated well under harsh conditions. The Liquicell extractor was run at a slight overpressure (about 2 %). This pressure was not accurately measured and appears the largest source of uncertainty in the analysis. The system gathered data every two minutes and two gas standards were used for a (linear calibration). Standards were run every hour. The first sample after the standard reads low because of the flow configuration and is omitted Columns PC_DATE= date PC_TIME = time YEAR_DAY = Julian day from PC computer (applies to XCO2 values) SHIP_JD= Julian day from ship SCS (applies to salinity and temperature values) DecLat = latitude DecLong= longitude xCO2(27.3)bowhound= xCO2 normalized to 27.3 using a relationship of 4.23 % per degree EQ_TEMP= temperature at the intake of the liquicell AdjLin_xCO2= measured xCO2 SCS_T= thermosalinograph temperature SCS_Sal = thermosalinograph salinity Sys2.6xCO2 = XCO2 from system in main lab Note there are no latitudes and longitude values for the last 1000 entries after the ship left the study area.