PSA file: C:\repeat\a16n_13\acq\Seasave.psa Date: 08/13/2013 Instrument configuration file: C:\repeat\a16n_13\acq\pri_24_2.xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : No Scan time added : Yes 1) Frequency 0, Temperature Serial number : 5403 Calibrated on : 18-Jun-13 G : 4.34811691e-003 H : 6.30354229e-004 I : 1.89818107e-005 J : 1.19581755e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 3338 Calibrated on : 10-May-13 G : -1.01957526e+001 H : 1.57743198e+000 I : -2.69477930e-003 J : 3.02939581e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 1035 Calibrated on : 07-Jan-11 C1 : -4.373825e+004 C2 : 4.277260e-001 C3 : 1.413200e-002 D1 : 3.420800e-002 D2 : 0.000000e+000 T1 : 2.988725e+001 T2 : -1.949980e-004 T3 : 4.187830e-006 T4 : 4.590370e-009 T5 : 0.000000e+000 Slope : 1.00000000 Offset : 0.00000 AD590M : 1.279900e-002 AD590B : -9.386546e+000 4) Frequency 3, Temperature, 2 Serial number : 2958 Calibrated on : 10-May-13 G : 4.39479263e-003 H : 6.72823388e-004 I : 2.99477848e-005 J : 2.77035344e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 3647 Calibrated on : 10-May-13 G : -1.00824661e+001 H : 1.61729969e+000 I : -1.78043023e-003 J : 2.30263771e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Oxygen, SBE 43 Serial number : 1666 Calibrated on : 18-Jun-13 Equation : Sea-Bird Soc : 5.56840e-001 Offset : -5.06400e-001 A : -3.60000e-003 B : 1.54370e-004 C : -3.07600e-006 E : 3.60000e-002 Tau20 : 1.25000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 7) A/D voltage 1, Oxygen, SBE 43, 2 Serial number : 1329 Calibrated on : 10-May-13 Equation : Sea-Bird Soc : 4.38650e-001 Offset : -5.00700e-001 A : -3.24640e-003 B : 1.74800e-004 C : -2.65680e-006 E : 3.60000e-002 Tau20 : 1.50000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 8) A/D voltage 2, Altimeter Serial number : 2 Calibrated on : Scale factor : 4.800 Offset : 0.000 9) A/D voltage 3, Free 10) A/D voltage 4, User Polynomial Serial number : 8756 Calibrated on : Sensor name : Metrox Load Cell [lbs] A0 : 0.00000000 A1 : 526.00000000 A2 : 0.00000000 A3 : 0.00000000 11) A/D voltage 5, Free 12) A/D voltage 6, User Polynomial, 2 Serial number : 507DR Calibrated on : Sensor name : Transmissometer [V] A0 : 0.00000000 A1 : 1.00000000 A2 : 0.00000000 A3 : 0.00000000 13) A/D voltage 7, User Polynomial, 3 Serial number : FLCDRTD-428 Calibrated on : Sensor name : Fluorometer [V] A0 : 0.00000000 A1 : 1.00000000 A2 : 0.00000000 A3 : 0.00000000 Scan length : 41 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: C:\repeat\a16n_13\data\rh10111.hex Timeout (seconds) at startup: 20 Timeout (seconds) between scans: 20 --------------------------------------------- Instrument port configuration: Port = COM5 Baud rate = 19200 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: SBE Carousel Number of bottles: 24 Port: COM1 Enable remote firing: NO Firing sequence: Sequential Tone for bottle fire confirmation uses PC sound card. --------------------------------------------- Header information: Header Choice = Include Default Header Information in File --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth and Average Sound Velocity Latitude when NMEA is not available: 0.00000000 Average Sound Velocity Minimum pressure [db]: 20.00000000 Minimum salinity [psu]: 20.00000000 Pressure window size [db]: 20.00000000 Time window size [s]: 60.00000000 Descent and Acceleration Window size [s]: 2.00000000 Plume Anomaly Theta-B: 0.00000000 Salinity-B 0.00000000 Theta-Z / Salinity-Z 0.00000000 Reference pressure [db] 0.00000000 Oxygen Window size [s]: 2.00000000 Apply hysteresis correction: 0 Apply Tau correction: 0 Potential Temperature Anomaly A0: 0.00000000 A1: 0.00000000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: NO --------------------------------------------- Mark Variables: Variables: Digits Variable Name [units] ------ --------------------- 0 Scan Count 4 Pressure, Digiquartz [db] 5 Temperature [ITS-90, deg C] 5 Temperature, 2 [ITS-90, deg C] 5 Salinity, Practical [PSU] 5 Salinity, Practical, 2 [PSU] 4 Oxygen, SBE 43 [umol/Kg] 4 Oxygen, SBE 43, 2 [umol/Kg] 5 Density [sigma-theta, Kg/m^3] 5 Density, 2 [sigma-theta, Kg/m^3] 6 Latitude [deg] 6 Longitude [deg] --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: YES XML wrapper and settings: YES Seconds between raw data updates: 0.00000000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC sound card. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO